Yet another lost guide wire


Department of Anesthesiology, Indira Gandhi Institute of Medical Sciences, Patna, Bihar, India

Correspondence Address:
Dr. Prakash K Dubey
E ¾, IGIMS Campus, Patna - 800 014, Bihar
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/sja.SJA_108_20

Rights and Permissions
Date of Submission05-Feb-2020
Date of Decision09-Feb-2020
Date of Acceptance09-Feb-2020
Date of Web Publication24-Sep-2020
 


How to cite this article:
Dubey PK, Rahul R, Bharti AK. Yet another lost guide wire. Saudi J Anaesth 2020;14:554-6

How to cite this URL:
Dubey PK, Rahul R, Bharti AK. Yet another lost guide wire. Saudi J Anaesth [serial online] 2020 [cited 2020 Nov 24];14:554-6. Available from: https://www.saudija.org/text.asp?2020/14/4/554/295989



Sir,

Central venous catheterization (CVC) is a routine technique used in the pediatric health care setup for various purposes. One of the rare complications of this procedure is the intravascular loss of a guidewire, which may lead to various life-threatening consequences. We present a case of a retained guidewire that was introduced through the femoral vein. Point of care ultrasonography was used during its retrieval.

A 6-month-old male child was referred to the emergency room from another hospital with a retained guidewire in the right femoral vein for 5 days. He was being treated for pneumonia and right femoral venous cannulation was attempted during management. Although he was carrying a radiological film showing the guidewire, apparently no attempt was made to retrieve it as they were waiting for the child to recover from his illness. Then he was referred to our center.

A fresh X-ray revealed a straight wire extending from the right groin to the right supraclavicular region [Figure 1]a. Interestingly, the tip of the guidewire was straight and not curved.
Figure 1: (a) X-ray showing the retained guidewire. (b) Guidewire (inset) being removed

Click here to view


The child was afebrile and all routine investigations were within normal limits. Keeping in view the position of the wire and the possible complications, it was decided to remove it under general anesthesia. After the institution of routine monitoring in the form of ECG, pulse oximeter and noninvasive arterial pressure, anesthesia was induced with fentanyl and propofol. A supraglottic airway was introduced following the administration of atracurium. Anesthesia was maintained with 2% sevoflurane and 50% oxygen in the air along with intermittent positive pressure ventilation.

It was decided to use point of care ultrasonography (M Turbo, Fujifilm SonoSite, Inc, Bothell, WA, USA) to locate the guidewire during retrieval to avoid further radiation exposure to the child. The guidewire tip was accurately visualized in the internal jugular vein and was retrieved surgically [Figure 1]b. On inspection, it was 23 cm in length and appeared to be a guidewire used for arterial cannulation. Neither of the ends was flexible or curved unlike the central venous catheter (CVC) guidewires available. The short length contributed to the intravascular loss of guidewire.

Neuromuscular blockade was reversed with neostigmine and atropine and the child made an uneventful recovery.

Intravascular loss of guidewire, although extremely rare, is a completely preventable complication. A retained guidewire results in costly investigations and procedures for its retrieval that are potentially harmful to the patient and may also lead to medicolegal issues.[1] Various modalities such as radiographs, CT scans, ultrasonography, echocardiography, and angiography have been used for the detection of the lost guidewires.[1] Our patient also had repeated radiation exposure and had to undergo surgery under general anesthesia that was completely avoidable.

A lost guidewire may remain asymptomatic and incidentally found on a routine X-ray done sometime later. The longest gap reported was 20 years while in one patient a retained guidewire was incidentally detected during the postmortem examination.[1]

The possible complications include arrhythmias, vascular damage, thromboembolic complications, infection, cardiac perforation, or tamponade.[2] Migration of a right femoral vein catheter guidewire into the circulation leading to damage of arch of aorta and cerebrum has also been reported.[3] The retained guidewires are also known to get fragmented inside the body.[1] Fortunately, there was no major complication in our patient in spite of the retention of a wire that did not have a flexible tip.

Although interventional radiology techniques are the preferred method for retrieval and removal, we preferred surgical removal as facilities for the former were not available at our center.[2] In spite of recovering from a major illness, the child tolerated the anesthetic procedure satisfactorily.

The risk factors identified for this complication include the insertion of catheters by trainees without adequate supervision, distractions during the procedure, and a high workload.[1] Many strategies have been suggested to avoid this complication. A quality improvement project to eliminate incidences of guidewire retention during CVC insertion has also been proposed.[4]

Important steps to avoid this include creating awareness of the possibility of retained guidewires and its complications; training on a mannikin; effective supervision; strong vigilance during insertion; and provision of an adequate workforce, especially outside routine hours, as well as avoiding insertion at night unless essential.[1]

Our case was especially different because apparently a guidewire from an arterial cannula was used for CVC insertion that was completely unwarranted. The rigid nature of this guidewire had the potential of causing serious damage. We believe that the simple step of maintaining a hold of the proximal tip of the wire during placement would have prevented this “never event.”

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Acknowledgements

The authors express their gratitude to Mr. Gopal Panday for his contribution in preparing the figure.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Pokharel K, Biswas BK, Tripathi M, Subedi A. Missed central venous guide wires: A systematic analysis of published case reports. Crit Care Med 2015;43:1745-56.  Back to cited text no. 1
    
2.
Srivastav R, Yadav V, Sharma D, Yadav V. Loss of guide wire: A lesson learnt review of literature. J Surg Tech Case Report 2013;5:78-81.  Back to cited text no. 2
[PUBMED]  [Full text]  
3.
Zhao S, Wang Z, Zhao Y. Loss of guidewire and its sequelae after central venous catheterization: A case report. Medicine (Baltimore) 2019;98:e16513.  Back to cited text no. 3
    
4.
Peh WM, Loh WJ, Phua GC, Loo CM. Eliminating guidewire retention during ultrasound guided central venous catheter insertion via an educational program, a modified CVC set, and a drape with reminder stickers. BMJ Quality Improvement Reports 2016; u209550.w3941 doi: 10.1136/bmjquality.u209550.w3941.  Back to cited text no. 4
    


    Figures

  [Figure 1]



 

Top
 
Previous article    Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  IN THIS Article
   References
   Article Figures

 Article Access Statistics
    Viewed125    
    Printed0    
    Emailed0    
    PDF Downloaded13    
    Comments [Add]    

Recommend this journal