Ultrasound-guided probe-generated artifacts stimulating ventricular tachycardia: A rare phenomenon
Rafat Shamim1, Rudrashish Haldar1, Ashutosh Kaushal2
1 Department of Anaesthesiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India 2 Department of Neuroanaesthesiology, All India Institute of Medical Sciences, New Delhi, India
Correspondence Address:
Rudrashish Haldar Department of Anaesthesiology, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Rae Bareilly Road, Lucknow, Uttar Pradesh India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/1658-354X.197340
|
|
Electrocardiographic (ECG) artifacts may arise due to interference, faulty earthing, and current leakages in biomedical equipment which might create clinical dilemmas in the perioperative settings. Piezoelectric signals generated by ultrasonography probe are another uncommon source which might be sensed by the ECG electrodes and produce tracings similar to pathological arrhythmias triggering false alarms and avoidable therapies. Anesthesiologists should be familiar with these uncommon sources which might produce these artifacts and they should be identified swiftly. |
|
|
|