Previous article
Next article |
|
ORIGINAL ARTICLE
Year : 2015 | Volume
: 9
| Issue : 2 | Page : 195-198
Influence of GlideScope assisted endotracheal intubation on intraocular pressure in ophthalmic patients
Nauman Ahmad1, Abdul Zahoor1, Waleed Riad2, Saeed Al Motowa3
1 Department of Anesthesia, King Khaled Eye Specialist Hospital, Riyadh, Saudi Arabia 2 Department of Anesthesia, St. Michael's Hospital, Toronto, Ontario, Canada 3 Department of Anterior Segment Division, King Khaled Eye Specialist Hospital, Riyadh, Saudi Arabia
Correspondence Address: Dr. Waleed Riad Department of Anesthesia, St. Michael's Hospital, 30 Bond Street, Toronto, Ontario M5B 1W8 Canada
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/1658-354X.152885
Clinical trial registration NCT01271621

Date of Web Publication | 10-Mar-2015 |
|
|

Background: Traditional Macintoch laryngoscopy is known to cause a rise in intraocular pressure (IOP), tachycardia and hypertension. These changes are not desirable in patients with glaucoma and open globe injury. GlideScope is a video laryngoscope that functions independent of the line of sight, reduces upward lifting forces for glottic exposure and requires less cervical neck movement for intubation, making it less stimulating than Macintosh laryngoscopy. Aim: The aim was to assess the variations in IOP and hemodynamic changes after GlideScope assisted intubation. Materials and Methods: After approval of the local Institutional Research and Ethical Board and informed patient consent, 50 adult American Society of Anesthesiologist I and II patients with normal IOP were enrolled in a prospective, randomized study for ophthalmic surgery requiring tracheal intubation. In all patients, trachea was intubated using either GlideScope or Macintoch laryngoscope. IOP of nonoperated eye, heart rate and blood pressure were measured as baseline, 1 min after induction, 1 min and 5 min after tracheal intubation. Results: IOP was not significantly different between groups before and after anesthetic induction and 5 min after tracheal intubation (P = 0.217, 0.726, and 0.110 respectively). The only significant difference in IOP was at 1 min after intubation (P = 0.041). No significant difference noted between groups in mean arterial pressure (P = 0.899, 0.62, 0.47, 0.82 respectively) and heart rate (P = 0.21, 0.72, 0.07, 0.29, respectively) at all measurements. Conclusion: GlideScope assisted tracheal intubation shown lesser rise in IOP at 1 min after intubation in comparison to Macintoch laryngoscope, suggesting that GlideScope may be preferable to Macintosh laryngoscope. Keywords: GlideScope, hemodynamic response, intraocular pressure, intubation
How to cite this article: Ahmad N, Zahoor A, Riad W, Al Motowa S. Influence of GlideScope assisted endotracheal intubation on intraocular pressure in ophthalmic patients. Saudi J Anaesth 2015;9:195-8 |
How to cite this URL: Ahmad N, Zahoor A, Riad W, Al Motowa S. Influence of GlideScope assisted endotracheal intubation on intraocular pressure in ophthalmic patients. Saudi J Anaesth [serial online] 2015 [cited 2022 May 24];9:195-8. Available from: https://www.saudija.org/text.asp?2015/9/2/195/152885 |
Introduction | |  |
Macintosh laryngoscopy has been traditionally used to facilitate tracheal intubation. This maneuver is associated with increased intraocular pressure (IOP), tachycardia and hypertension. [1]
These changes are not suitable for patients with glaucoma and open globe injury. Conventional intubation by Macintosh laryngoscope need to bring Oropharyngeal and laryngeal axis into line of sight that can develop stress response in patients. GlideScope video laryngoscope functions independently of the line of sight, reduces upward lifting forces to expose the glottis and requires less cervical neck movement for intubation, making it potentially less stimulating than Macintosh laryngoscope. [2] There is also a significant increase in the percentage of glottic opening visibility when using the GlideScope with cervical spine immobilization. [3]
To the best of our knowledge changes in IOP in ophthalmic patients while preforming endotracheal intubation using Glidoscope have not been studied. The aim of this study was to compare the variations in IOP and hemodynamics, associated with GlideScope versus conventional Macintosh laryngoscope assisted intubation.
Materials and Methods | |  |
We obtained approval of Institutional Research and Ethical Board and informed written patient consent. This was a prospective, randomized study for ophthalmic surgery requiring tracheal intubation. Power analysis indicated that 25 patients were required in each arm to detect a difference of 30% in the IOP with 80% power and alpha error was set to 0.050 two-sided. Fifty adults patients with normal IOP and having American Society of Anesthesiologist (ASA) I and II were enrolled. Any patients with the previous history of raised IOP, cardiovascular hypertensive disease or renal, respiratory and neurological involvement were excluded from the study.
During preoperative visits all patients were assessed by a co-investigator for difficulty in laryngoscopy or visualization of the glottis by Mallampati grading, measuring thyromental distance and extension at the atlanto-occipital joint. Patients with predicted difficult intubation were excluded from this study. All patients were premedicated with midazolam (0.1 mg/kg) 1 h before surgery. Usual monitoring was applied to all patients. General anesthesia was induced with propofol, followed by rocuronium in a standardized manner in all patients. Patients were randomly assigned to either GlideScope (group 1) or Macintosh laryngoscope (group 2) for intubation. All intubations were performed by a single anesthesiologist who was experienced in both techniques. Baseline IOP (nonoperated eye), heart rate, mean blood pressure and duration of intubation were measured 1 min after induction, followed by 1 and 5 min after tracheal intubation.
Data analysis was performed using Statistical Package for Social Studies (SPSS 19) (IBM, Boston, USA). Differences between the two groups with regard to intubation time, IOP, heart rate and mean arterial pressure (MAP) were tested by Mann-Whitney U-test. P < 0.05 were considered as statistically significant.
Results | |  |
Demographic and clinical data were similar for age, sex, weight, height, Mallampati/ASA classifications and duration of surgery among both groups. All tracheal intubations were successful at first attempt. GlideScope, generally provided a laryngoscopic view equal or better than that of direct laryngoscopy. IOP was not significantly different between groups before and after induction and 5 min after tracheal intubation (P = 0.217, 0.726, 0.110 respectively). The only significant difference was lesser rise in IOP at 1 min after intubation (P = 0.041) in GlideScope group in comparison with Macintosh group. No significant difference was found between the groups in MAP (P = 0.899, 0.62, 0.47, 0.82 respectively) and in the heart rate (P = 0.21, 0.72, 0.07, 0.29 respectively) at all measurements. Duration of intubation was slightly longer in group 1 in comparison to group 2 (20.12 ± 8.05 and 16.12 ± 5.67 s respectively), but was not significant statistically (P = 0.079) [Table 1] and [Figure 1], [Figure 2] and [Figure 3]. | Figure 1: Intraocular pressure in nonoperated eye after GlideScope assisted endotracheal intubation versus conventional method
Click here to view |
 | Figure 2: Mean arterial blood pressure after GlideScope assisted endotracheal intubation versus conventional method
Click here to view |
 | Figure 3: Heart rate in after GlideScope assisted endotracheal intubation versus conventional method
Click here to view |
Discussion | |  |
Laryngoscopy and endotracheal intubation is likely to produce tachycardia and hypertension and an increase in IOP. [4] IOP is elevated secondary to increased sympathetic activity that causes vasoconstriction leading to a rise in central venous pressure, which determines episcleral venous pressure. A raised episcleral venous pressure could lead to vitreous chamber venous engorgement and a reduction in aqueous humour drainage, both of which lead to a rise in IOP. Therefore, prevention from acute rise in IOP becomes important during laryngoscopy and endotracheal intubation in patients with acute glaucoma and open globe injuries. An acutely raised IOP may cause expulsion of the globe contents through a surgical wound or traumatic perforation, or may lead to retinal artery occlusion with change from critical disc perfusion into disc ischemia. [5]
Intubation with GlideScope provides a clear view of glottis without alignment of the oral, pharyngeal, and laryngeal axes and less manipulation of airway, which may cause less stimulation of the sympathetic system and resulting in less increase in the IOP and other hemodynamic parameters.
Different approaches have been used to limit the intubation responses and a rise in IOP associated with endotracheal intubation. Brain's laryngeal mask airway and intubating laryngeal mask airway have shown a clear advantage over direct laryngoscopy for tracheal intubation in minimizing increase in IOP with lesser sympathetic stimulation than endotracheal tube insertion. [1],[6] Takahashi et al. [7] compared conventional laryngoscope with Lightwand intubating device (Trachlight) for hemodynamic changes, but could not find any difference in response. On contrary, Kihara et al. [8] has shown that intubating laryngeal mask airway Fastrach and the Trachlight Lightwand attenuated the hemodynamic stress response to tracheal intubation in comparison with Macintosh laryngoscope in hypertensive patients. Suresh et al. [9] concluded that McCoy laryngoscope in comparison to Macintosh laryngoscope resulted in significantly less rise in IOP and clinically less marked increase in hemodynamic response to laryngoscopy and intubation. This is also similar to a reduction in hemodynamic responses to tracheal intubation by the Bonfils retromolar fiberscope. [10]
In another study by Turkstra C-spine motion was reduced 50% at the C2-5 segment using the GlideScope. [11] Li et al. [12] showed that the hemodynamic responses caused by nasotracheal intubation were more severe than Fiber optic bronchoscope, followed by Macintosh direct laryngoscope, and least to GlideScope video-laryngoscope. In a recent report Pournajafian et al. [13] noted no significant variations in hemodynamic values while comparing between the GlideScope video laryngoscope and Macintosh laryngoscope during orotracheal intubation.
Most of the above mentioned studies have measured the hemodynamic responses instead of IOP using different tools for intubation but our study is unique in a way that we measured the direct changes in IOP when comparing GlideScope with conventional Macintosh laryngoscope. The baseline IOP was similar in both groups in our study. IOP significantly decreased in both the groups following induction (preintubation), [Figure 1], which was more likely due to the effect of anesthetic agents [14] but 1 min after intubation, GlideScope assisted tracheal intubation caused less rise in the IOP in comparison to Macintoch laryngoscope. GlideScope videolaryngoscopy took slightly longer time to intubate the trachea compared with the Macintosh blade despite the improved glottic view. This finding is well reported in the literature. [13],[15] All tracheal intubations were successful on the first attempt. In most patients, the GlideScope provided a laryngoscopic view equal to or better than that of direct laryngoscopy. It has potential advantages over standard direct laryngoscopy for difficult intubations. [16]
Conclusion | |  |
GlideScope may be preferable for use in ophthalmic patients in whom a rise in IOP is undesirable. However, further clinical trials are needed in patients with glaucoma to further evaluate the benefits of GlideScope.
Acknowledgments | |  |
The authors acknowledge the efforts of Rajiv Khandekar, MD and Nasira Asghar, MS in assisting the statistical analysis.
References | |  |
1. | Bukhari SA, Naqash I, Zargar J, Nengroo S, Mir AW. Pressor responses and intraocular pressure changes following insertion of laryngeal mask airway: Comparison with tracheal tube insertion. Indian J Anaesth 2003;47:473-5. |
2. | Tsai P, Chen B. Hemodynamic responses to endotracheal intubation comparing the airway scope ® , glidescope ® , and macintosh laryngoscopes. Internet J Anesthesiol 2009;24:2. |
3. | Kim HJ, Chung SP, Park IC, Cho J, Lee HS, Park YS. Comparison of the GlideScope video laryngoscope and Macintosh laryngoscope in simulated tracheal intubation scenarios. Emerg Med J 2008;25:279-82. |
4. | Watcha MF, White PF, Tychsen L, Stevens JL. Comparative effects of laryngeal mask airway and endotracheal tube insertion on intraocular pressure in children. Anesth Analg 1992;75:355-60. |
5. | Murgatroyd H, Bembridge J. Intraocular pressure. Contin Educ Anaesth Crit Care Pain 2008;8:100-3. |
6. | Bharti N, Mohanty B, Bithal PK, Dash M, Dash HH. Intra-ocular pressure changes associated with intubation with the intubating laryngeal mask airway compared with conventional laryngoscopy. Anaesth Intensive Care 2008;36:431-5. |
7. | Takahashi S, Mizutani T, Miyabe M, Toyooka H. Hemodynamic responses to tracheal intubation with laryngoscope versus lightwand intubating device (Trachlight) in adults with normal airway. Anesth Analg 2002;95:480-4. |
8. | Kihara S, Brimacombe J, Yaguchi Y, Watanabe S, Taguchi N, Komatsuzaki T. Hemodynamic responses among three tracheal intubation devices in normotensive and hypertensive patients. Anesth Analg 2003;96:890-5. |
9. | Suresh S, Karampal S, Naresh S, Punam R. Intraocular pressure changes following laryngoscopy and intubation: McCoy versus Macintosh laryngoscope. Sri Lankan J Anaesthesiol 2012;20:73-7. |
10. | Boker A, Almarakbi WA, Arab AA, Almazrooa AA. Reduced hemodynamic responses to tracheal intubation by the Bonfils retromolar fiberscope: A randomized controlled study. Middle East J Anaesthesiol 2011;21:385-90. |
11. | Turkstra TP, Craen RA, Pelz DM, Gelb AW. Cervical spine motion: A fluoroscopic comparison during intubation with lighted stylet, GlideScope, and Macintosh laryngoscope. Anesth Analg 2005;101:910-5. |
12. | Li XY, Xue FS, Sun L, Xu YC, Liu Y, Zhang GH, et al. Comparison of hemodynamic responses to nasotracheal intubations with GlideScope video-laryngoscope, Macintosh direct laryngoscope, and fiberoptic bronchoscope. Zhongguo Yi Xue Ke Xue Yuan Xue Bao 2007;29:117-23. |
13. | Pournajafian AR, Ghodraty MR, Faiz SH, Rahimzadeh P, Goodarzynejad H, Dogmehchi E. Comparing glidescope video laryngoscope and macintosh laryngoscope regarding hemodynamic responses during orotracheal intubation: A randomized controlled trial. Iran Red Crescent Med J 2014;16:e12334. |
14. | Mirakhur RK, Shepherd WF, Elliott P. Intraocular pressure changes during rapid sequence induction of anesthesia: Comparison of propofol and thiopentone in combination with vecuronium. Br J Anaesth 1988;60:379-83. |
15. | Healy DW, Picton P, Morris M, Turner C. Comparison of the glidescope, CMAC, storz DCI with the Macintosh laryngoscope during simulated difficult laryngoscopy: A manikin study. BMC Anesthesiol 2012;12:11. |
16. | Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Moult M. The glidescope video laryngoscope: Randomized clinical trial in 200 patients. Br J Anaesth 2005;94:381-4. |
[Figure 1], [Figure 2], [Figure 3]
[Table 1]
This article has been cited by | 1 |
Videolaryngoscopy versus direct laryngoscopy for adults undergoing tracheal intubation |
|
| Jan Hansel, Andrew M Rogers, Sharon R Lewis, Tim M Cook, Andrew F Smith | | Cochrane Database of Systematic Reviews. 2022; 2022(4) | | [Pubmed] | [DOI] | | 2 |
Comparison of the effects of the McGRATH MAC, C-MAC, and Macintosh laryngoscopes on the intraocular pressures of non-ophthalmic patients: A prospective, randomised, clinical trial |
|
| Ahmet Selim Ozkan, Sedat Akbas, Erol Karaaslan, Nihat Polat | | Anaesthesia Critical Care & Pain Medicine. 2021; 40(6): 100974 | | [Pubmed] | [DOI] | | 3 |
Use of the GlideScope does not lower the hemodynamic response to tracheal intubation more than the Macintosh laryngoscope: a systematic review and meta-analysis |
|
| Hiroshi Hoshijima, Koichi Maruyama, Takahiro Mihara, Aiji Sato Boku, Toshiya Shiga, Hiroshi Nagasaka | | Medicine. 2020; 99(48): e23345 | | [Pubmed] | [DOI] | | 4 |
COMPARISON OF INTRAOCULAR PRESSURE CHANGES IN PATIENTS UNDERGOING CATARACT SURGERY, BASED ON LOCAL AND GENERAL ANAESTHESIA WITH LARYNGEAL MASK AIRWAY AND LARYNGOSCOPY TYPES |
|
| Asma Amiri Domari, Ali Sarkoohi, Yousef Taghavi, Sayed Hamid Pakzad, Hamid Bakhshi | | Journal of Evolution of Medical and Dental Sciences. 2019; 8(22): 1762 | | [Pubmed] | [DOI] | | 5 |
Videolaryngoscopy versus direct laryngoscopy for adult patients requiring tracheal intubation |
|
| Sharon R Lewis, Andrew R Butler, Joshua Parker, Tim M Cook, Andrew F Smith | | Cochrane Database of Systematic Reviews. 2016; 2016(12) | | [Pubmed] | [DOI] | |
|
 |
|
|