CASE REPORT
Year : 2013  |  Volume : 7  |  Issue : 4  |  Page : 474-475

Henna dye: A cause of erroneous pulse oximetry readings


Department of Anesthesiology, Military Hospital Med V Rabat, University of Med V Souissi Rabat, Morocco

Correspondence Address:
Mustapha Bensghir
Post Box 8840, Rabat Agdal, Rabat
Morocco
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.121052

Rights and Permissions
Date of Web Publication7-Nov-2013
 

  Abstract 

Pulse oximetry is a noninvasive and continuous monitoring of the pulsed saturation of hemoglobin oxygen. Because of its simplicity and usefulness, it is part of monitoring recommended for any anesthesia in the operating room. Different factors may limit the use of this monitoring. We report a case of difficulty monitoring by the presence of henna in a patient scheduled for general anesthesia in prone position.

Keywords: Anesthesia, henna dye, pulse oximetry, surgery


How to cite this article:
Bensghir M, Houba A, El Hila J, Ahtil R, Azendour H, Kamili ND. Henna dye: A cause of erroneous pulse oximetry readings. Saudi J Anaesth 2013;7:474-5

How to cite this URL:
Bensghir M, Houba A, El Hila J, Ahtil R, Azendour H, Kamili ND. Henna dye: A cause of erroneous pulse oximetry readings. Saudi J Anaesth [serial online] 2013 [cited 2023 Mar 27];7:474-5. Available from: https://www.saudija.org/text.asp?2013/7/4/474/121052


  Introduction Top


Pulse oximetry is a noninvasive, continuous monitoring of the saturation of oxygen when bound to hemoglobin in the blood. Although its usefulness cannot be doubted, the technique has a number of limitations, which may lead to inaccurate readings. [1] The following case report focuses on one such factor which can alter the accuracy of the pulse oximeter.


  Case report Top


A 65-year-old female presented in hospital for a herniated disc surgery. A preoperative clinical exam showed a darkened skin coloration of the hands and toes.

A week prior to admission, she had applied henna on her hands and feet. On the day of the surgery, she was shifted to the operating room for the procedure under general anesthesia. Preinduction monitors using pulse oximeter failed to produce a waveform on reading from any of the fingers or toes. Equipment failure was excluded by trying the probe on members of the staff. A change of oximeter also failed to produce a signal. The alternative of using others sites like the earlobe for monitoring was not viable because of the prone position required the surgery. The procedure was then delayed to a later date. Repeated attempts to clean fingers with a diluted bleach solvent did help in mitigating the dark coloration of the skin of the fingers [Figure 1]. The patient could be monitored successfully by applying the probe to the fingers four days later.
Figure 1: skin fingers pigmentation after several attempts using solvent

Click here to view



  Discussion Top


A pulse oximeter functions by evaluating the transmission of light at two wavelengths. Any interference in the transmission of light is likely to affect the oxygen saturation readings. One of the factors which may interfere with pulse oximeter readings is the dark skin pigmentation caused by certain skin dyes. [2],[3]

Henna (Lawsonia inermis) is a member of the family Lythracae, and is a commonly used dye. Henna, when newly applied, causes a black discolouration of the skin. During this initial phase, it absorbs all wavelengths of visible light, allowing only infrared rays to pass through. This results in the failure to monitor oxygen saturation using pulse oximeters. [3] Unlike other colorants, henna penetrates the superficial layers of the skin and cannot be removed immediately by common solvents. [4] In our case, the patient's use of solvent has helped to mitigate the black color of the skin to some extent, giving it a reddish color.

Some researchers have investigated the effect of henna on the accuracy of pulse oximetry. They found that black henna caused errors in oxygen saturation readings while red henna did not. [5] Another study showing that the style of the henna stain can affect the readings of the pulse oximeter. Patients who had henna paste applied to the thumbnails only showed an insignificant effect on oximetry but those in whom the entire digit was stained were at significantly higher risk for inaccurate readings when red henna was used. Our case report illustrates the difficulty of using pulse oximetry in patients who have used henna, especially in context of our culture where is used often. It can affect the accuracy of pulse oximetry readings. We suggest advising patients scheduled for programmed surgery, not to apply henna on their hands, two weeks prior to scheduled interventions.


  Acknowledgment Top


The author thanks YD and YB for their support.

 
  References Top

1.Stoneham MD. Uses and limitations of pulse oximetry. Br J Hosp Med 1995;54:35-41.  Back to cited text no. 1
    
2.Valdez-Lowe C, Ghareeb SA, Artinian NT. Pulse oximetry in adults. Am J Nurs 2009;109:52-9.  Back to cited text no. 2
    
3.Hameedullah, Rauf MA, Khan FA. Henna Paste and Pulse oximetry: Effect of different methods of application. J Anaesth Clin Pharmacol 2002;18:193-6.  Back to cited text no. 3
    
4.Samman YS, Rahimi JS, Wali SO, Krayem AB, Abdelaziz MM. Effects of henna dye on oxygen saturation reading using pulse oximetry. Saudi Med J 2006;27:268-9 .  Back to cited text no. 4
    
5. AL-Majed SA, Harakati MS. The effect of henna paste on oxygen saturation reading obtained by pulse oximetry. Trop Geogr Med 1994;46:38-9.  Back to cited text no. 5
    


    Figures

  [Figure 1]


This article has been cited by
1 Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis
Ana M. Cabanas, Macarena Fuentes-Guajardo, Katina Latorre, Dayneri León, Pilar Martín-Escudero
Sensors. 2022; 22(9): 3402
[Pubmed] | [DOI]
2 Esmaltes de uñas ecológicos y medición de la saturación periférica de oxígeno en pacientes quirúrgicos. Presentación de casos
Yosleiny Díaz-Reyes, Ayde María Pérez-Quiala, Jorge Humberto Díaz-Rodríguez
Revista Mexicana de Anestesiología. 2022; 45(4): 285
[Pubmed] | [DOI]
3 The Effect of Nail Polish and Henna on the Measures of Pulse Oximeters in Healthy Persons
Sevgi Deniz Dogan, Seda Karaçay Yikar, Sevban Arslan, Evsen Nazik
Journal of PeriAnesthesia Nursing. 2021; 36(5): 532
[Pubmed] | [DOI]
4 Importance and use of pulse oximeter in COVID-19 pandemic: general factors affecting the sensitivity of pulse oximeter
Kirtikumar C. Badgujar, Ashish B. Badgujar, Dipak V. Dhangar, Vivek C. Badgujar
Indian Chemical Engineer. 2020; 62(4): 374
[Pubmed] | [DOI]
5 Difficult peripheral venous cannulation in a patient with henna tattoo
SHEEMA SIRAJ
British Journal of Anaesthesia. 2016; 117: el_14075
[Pubmed] | [DOI]



 

Top
 
Previous article    Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  IN THIS Article
   Abstract
  Introduction
  Case report
  Discussion
  Acknowledgment
   References
   Article Figures

 Article Access Statistics
    Viewed2725    
    Printed60    
    Emailed0    
    PDF Downloaded177    
    Comments [Add]    
    Cited by others 5    

Recommend this journal