An easy solution to obstructed sampling line: Auxillary oxygen flowmeter


Department of Anaesthesiology and Critical Care, Pt. BDS PGIMS, Rohtak, Haryana, India

Correspondence Address:
Swati Chhabra
95, Green Park, Hisar 125001, Haryana
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.101239

Rights and Permissions
Date of Web Publication21-Sep-2012
 


How to cite this article:
Chhabra S, Taxak S. An easy solution to obstructed sampling line: Auxillary oxygen flowmeter. Saudi J Anaesth 2012;6:312-3

How to cite this URL:
Chhabra S, Taxak S. An easy solution to obstructed sampling line: Auxillary oxygen flowmeter. Saudi J Anaesth [serial online] 2012 [cited 2023 Mar 30];6:312-3. Available from: https://www.saudija.org/text.asp?2012/6/3/312/101239

Sir,

This is with reference to the correspondence by Nataraj Madagondapalli Srinivasan, "Use of heat and moisture exchanger in intubated patients reduces blockage in gas sampling tube of the mass spectrometer," [1] in the Jan-Mar 2011 issue of the Saudi Journal of Anaesthesia. We read it with interest as it discusses a rather common problem.

Capnography is being widely used intraoperatively for confirmation of endotracheal tube placement and for adequacy of ventilation during anesthesia. If facilities are available, the gas sampled for capnography is also used for anesthetic gas monitoring using the principle of mass spectrometry.

Obstruction of sampling line due to water droplets or secretions [2] is not infrequent, especially during long procedures or when a single sampling line is used for a number of cases. Although the water trap prevents entry of these droplets into the monitor, tubing obstruction can lead to erroneous readings or no readings at all.

Many methods are described in the literature to get rid of the obstruction, viz. high gas flows, flushing with a syringe and heat and moisture exchanger. [1]

Recently, during a procedure over the head and neck, we encountered an obstruction such that our anesthesia gas monitor could not analyze the sampled gas to give us readings. The distal end of the sampling line was under the drapes and thus not accessible. The proximal end was not compatible for syringe attachment [Figure 1]. Therefore, we connected the proximal end to the auxillary oxygen flowmeter of the anesthesia machine with oxygen flow at 7 L/min to flush out the obstruction [Figure 2].
Figure 1: Incompatibility of syringe and sampling line

Click here to view
Figure 2: Proximal end of sampling line connected to auxillary oxygen flowmeter

Click here to view


When the sampling line was reattached to the monitor [Figure 3], proper capnograph and readings were obtained. Thus, this becomes an effective method of relieving the sampling line obstruction, which does not require any additional equipment. Although there is a disadvantage of dilution of anesthetic gases, it is very transient as high oxygen flows are used for a short duration of time.
Figure 3: AGM with water trap and proximal sampling line

Click here to view


 
  References Top

1.Srinivasan NM. Use of heat and moisture exchanger in intubated patients reduces the blockage in gas sampling tube of the mass spectrometer. Saudi J Anaesth 2011;5:112-3.  Back to cited text no. 1
[PUBMED]  Medknow Journal  
2.Skeehan TM, Biebuyck JF. Erroneous mass spectrometer data caused by a faulty patient sampling tube: case report and laboratory study. J Clin Monit 1991;7:313-9.  Back to cited text no. 2
[PUBMED]    


    Figures

  [Figure 1], [Figure 2], [Figure 3]


This article has been cited by
1 An indigenous method to prevent intraoperative kinking of gas sampling line
LeninBabu Elakkumanan, Sivaraman Baskaran, Senthilnathan Muthapillai
Indian Journal of Anaesthesia. 2013; 57(6): 631
[Pubmed] | [DOI]



 

Top
 
Previous article    Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  IN THIS Article
   References
   Article Figures

 Article Access Statistics
    Viewed2469    
    Printed80    
    Emailed0    
    PDF Downloaded129    
    Comments [Add]    
    Cited by others 1    

Recommend this journal