ORIGINAL ARTICLE
Year : 2017  |  Volume : 11  |  Issue : 1  |  Page : 15-20

Tissue type determination by impedance measurement: A bipolar and monopolar comparison


1 Wolfson School of Mechanical and Manufacturing Engineering, Loughborough University, Loughborough, Leicestershire LE11 3TU, UK
2 Department of Anaesthesia, University Hospitals of Leicester NHS Trust, Leicester, Leicestershire LE1 5WW, UK

Correspondence Address:
Atul Gaur
Department of Anaesthesia, University Hospitals of Leicester NHS Trust, Leicester, Leicestershire LE1 5WW
UK
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.197334

Rights and Permissions

Background: In certain medical applications, it is necessary to be able to determine the position of a needle inside the body, specifically with regards to identifying certain tissue types. By measuring the electrical impedance of specific tissue types, it is possible to determine the type of tissue the tip of the needle (or probe) is at. Materials and Methods: Two methods have been investigated for electric impedance detection; bipolar and monopolar. Commercially available needle electrodes are of a monopolar type. Although many patents exist on the bipolar setups, these have not as yet been commercialized. This paper reports a comparison of monopolar and bipolar setups for tissue type determination. In vitro experiments were carried out on pork to compare this investigation with other investigations in this field. Results: The results show that both monopolar and bipolar setups are capable of determining tissue type. However, the bipolar setup showed slightly better results; the difference between the different soft tissue type impedances was greater compared to the monopolar method. Conclusion: Both monopolar and bipolar electrical impedance setups work very similarly in inhomogeneous volumes such as biological tissue. There is a clear potential for clinical applications with impedance-based needle guidance, with both the monopolar and bipolar setups. It is, however, worth noting that the bipolar setup is more versatile.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed2213    
    Printed20    
    Emailed0    
    PDF Downloaded81    
    Comments [Add]    

Recommend this journal