LETTER TO EDITOR
Year : 2014  |  Volume : 8  |  Issue : 4  |  Page : 567-568

An unexpected error in oxygen humidifier


1 Department of Anaesthesiology, BLDE University and Research Centre, Bijapur, Karnataka, India
2 Department of Anaesthesiology, Bidar Medical Sciences, Bidar, Karnataka, India
3 Department of Anaesthesiology, Krishna Institute of Medical Sciences, Karad, Maharashtra, India

Correspondence Address:
Dr. Akshaya N Shetti
Department of Anaesthesiology, Krishna Institute of Medical Sciences, Karad, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.140911

Rights and Permissions
Date of Web Publication16-Sep-2014
 


How to cite this article:
Karigar SL, Kunakeri S, Shetti AN. An unexpected error in oxygen humidifier. Saudi J Anaesth 2014;8:567-8

How to cite this URL:
Karigar SL, Kunakeri S, Shetti AN. An unexpected error in oxygen humidifier. Saudi J Anaesth [serial online] 2014 [cited 2019 Dec 10];8:567-8. Available from: http://www.saudija.org/text.asp?2014/8/4/567/140911

Sir,

Oxygen is an important and essential gas for all living beings. Many different techniques are used to deliver oxygen to patients either to treat pathological conditions or to supplement during the perioperative period. [1] Equipment malfunction is a problem of particular importance during anesthesia and resuscitation. [2] Such problems hence extend from operation theater to high dependency unit (HDU) if care is not taken. Oxygen can be delivered in dry or humidified form. Advantages of supplementation of humidified oxygen are that, it prevents hypothermia, inflammation of the airway epithelium, bronchospasm, atelectasis and airway obstruction. [3] Humidified oxygen is one of the methods of delivery and commonly used in perioperative area and in HDU. Here, we share our experience of patients in the post-operative area who failed to maintain oxygen saturation which were actually preventable. Two times we observed fall in saturation in our post-operative setup in two patients one who underwent thoracotomy and other cardiac surgery. After extubation patients were conscious, oriented and pattern of breathing was acceptable. After the supplementation of oxygen at 10 L/min through simple mask, there was no improvement in saturation instead we noticed a drop in saturation from 96 (with room air) to 94% in both patients. After thorough inspection of humidifier we could see bubbles in humidifying chamber with bobbin positioned at 10 L marking with the absence of gas at output. On detailed inspection, the lid of distilled water chamber was damaged [Figure 1] in first and the leak was appreciated at the junction of the lid and input port of oxygen into the humidifying chamber [Figure 2] in the second case. The fall in saturation was due to the loss of humidified oxygen through low resistance area, i.e., through a hole in damaged lid in first and at the leaking site in the second scenario. Hence patients were actually not supplemented with humidified oxygen. Visual evidence of bubble formation in humidifying chamber and bobbin position at the set level may confirm entry of oxygen in humidifying chamber but doesn't assure delivery of oxygen to the patient. Hence, it is important to have leak proof and undamaged oxygen humidifier. The key message here is one should not jump into conclusion which may lead to invasive techniques like mechanical ventilation if saturation is not maintained, rather technical or mechanical errors should be identified and rectified. Inadequate experience and insufficient familiarity with equipment does contribute for such failure. High index of suspicion is necessary to identify such errors in the oxygen delivery devices which are most commonly practiced. It is important to have good quality medical devices and should never be a compromise in emergency and HDU.
Figure 1: Arrow showing damaged part in the lid

Click here to view
Figure 2: Leak at the junction

Click here to view


 
  References Top

1.St Clair N, Touch SM, Greenspan JS. Supplemental oxygen delivery to the nonventilated neonate. Neonatal Netw 2001;20:39-46.  Back to cited text no. 1
    
2.Cuquemelle E, Lellouche F. Assessment of humidification performance: Still no easy method! Respir Care 2013;58:1559-61.  Back to cited text no. 2
    
3.McIntyre JW. Anesthesia equipment malfunction: Origins and clinical recognition. Can Med Assoc J 1979;120:931-4.  Back to cited text no. 3
[PUBMED]    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
Previous article    Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  IN THIS Article
   References
   Article Figures

 Article Access Statistics
    Viewed892    
    Printed12    
    Emailed0    
    PDF Downloaded65    
    Comments [Add]    

Recommend this journal