ORIGINAL ARTICLE
Year : 2010  |  Volume : 4  |  Issue : 3  |  Page : 131-136

Comparative study between I-gel, a new supraglottic airway device, and classical laryngeal mask airway in anesthetized spontaneously ventilated patients


Department of Anesthesia and Intensive Care, Faculty of Medicine, Suez Canal University, Ismailia, Egypt

Correspondence Address:
Hossam M Atef
Department of Anesthesia and Intensive Care, Faculty of Medicine, Suez Canal University, Ismailia
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.71250

Rights and Permissions

Objective: To compare two different supraglottic airway devices, the laryngeal mask airway (LMA) and the I-gel, regarding easiness of insertion of the device, leak pressure, gastric insufflation, end tidal CO 2 , oxygen saturation, hemodynamic and postoperative complications in anesthetized, spontaneously ventilated adult patients performing different non-emergency surgical procedures. Materials and Methods: The study was carried out as a prospective, randomized, clinical trial among 80 patients who underwent different surgical procedures under general anesthesia with spontaneous ventilation in supine position. They were equally randomized into two groups: I-gel and LMA groups. Both the devices were compared with regard to heart rate, arterial BP, SPO 2 , end-tidal CO 2 , number and duration of insertion attempts, incidence of gastric insufflation, leak pressure and airway assessment after removal of the device. Results: No statistically significant difference was reported between both the groups, regarding heart rate, arterial BP, SPO 2 and end-tidal CO 2 . The mean duration of insertion attempts was 15.6±4.9 seconds in the I-gel group, while it was 26.2±17.7 seconds in the LMA group. The difference between both the groups regarding duration of insertion attempts was statistically significant (P=0.0023*), while the number of insertion attempts was statistically insignificant between both the study groups (P>0.05). Leak pressure was (25.6±4.9 vs. 21.2±7.7 cm H 2 O) significantly higher among studied patients of the I-gel group (P=0.016*) and the incidence of gastric insufflation was significantly more with LMA group 9 (22.5%) vs. I-gel group (5%) (P=0.016). Conclusion: Both LMA and I-gel do not cause any significant alteration in the hemodynamic status of the patients, end tidal CO 2 , and SPO 2 . The postoperative complications were not significantly different except nusea and vomiting was statistically significant higher in LMA group (P=0.032). among both LMA and I-gel patients. Insertion of I-gel was significantly easier and more rapid than insertion of LMA. Leak pressure was significantly higher with I-gel than LMA and thus incidence of gastric insufflation was significantly lower with I-gel.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed4135    
    Printed181    
    Emailed3    
    PDF Downloaded461    
    Comments [Add]    
    Cited by others 6    

Recommend this journal