CASE REPORT
Year : 2010  |  Volume : 4  |  Issue : 2  |  Page : 108-110

Endotracheal tube defects: Hidden causes of airway obstruction


Department of Anesthesia, King Abdul-Aziz Medical City, King Fahad National Guard Hospital, Riyadh, Saudi Arabia

Correspondence Address:
Khalid Sofi
Department of Anaesthesia, MC 1311, King Abdulaziz Medical City, King Fahad National Guard Hospital, P.O Box 22490, Riyadh 11426
Saudi Arabia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1658-354X.65123

Rights and Permissions
Date of Web Publication6-Jul-2010
 

  Abstract 

Manufacturing defects of endotracheal tube (ETT) are still encountered in anesthesia practice. Many such defects go unnoticed during routine inspection prior to their use. Such defects in ETT may lead to partial or complete airway obstruction in an intubated patient. We report a case of partial airway obstruction with a prepacked, single use, uncuffed ETT due to a manufacturing defect in the form of a plastic meniscus at the distal end of the tube. This case report highlights the significance of standard monitoring of ventilation and the role of a vigilant clinician in detecting such defects in avoiding critical events as can arise from the use of such defective ETTs. It also emphasizes the need for double checking ETTs prior to their use.

Keywords: Endotracheal tube, ventilation, monitoring, intubation


How to cite this article:
Sofi K, El-Gammal K. Endotracheal tube defects: Hidden causes of airway obstruction. Saudi J Anaesth 2010;4:108-10

How to cite this URL:
Sofi K, El-Gammal K. Endotracheal tube defects: Hidden causes of airway obstruction. Saudi J Anaesth [serial online] 2010 [cited 2019 Oct 19];4:108-10. Available from: http://www.saudija.org/text.asp?2010/4/2/108/65123


  Introduction Top


Difficulty in ventilating a tracheally intubated patient, as a result of endotracheal tube (ETT) defects, is known in anesthesia practice. Anesthesia gas delivery malfunction, obstruction of breathing circuit and poor pulmonary compliance are common causes which can make ventilation in an intubated patient difficult. Despite the common practice of visual inspection and testing of ETTs for physical defects prior to use, some manufacturing defects still go unnoticed. [1] Obstruction of an ETT is a potentially life-threatening event. Obstruction by mucus, blood or kinking is not uncommon; however, obstruction from a structural defect is also not very rare. [2] When difficulty in ventilation through an ETT is encountered, a quick differential diagnosis is warranted, which includes bronchospasm, pneumothorax, chest wall rigidity and equipment malfunction. We report a case of partial airway obstruction with a 4.5-mm ID uncuffed ETT, due to a meniscus at the distal end of the tube, which was missed on routine inspection prior to use.


  Case Report Top


A 2-year-old boy weighing 12 kg, ASA 1, with previous urethral injury, was scheduled for an elective urethroplasty. The anesthesia plan included general anesthesia with endotracheal intubation and single shot caudal epidural. The child was premedicated with 5 mg of oral midazolam 1 hour prior to induction. After establishing basic monitoring -Electrocardiography (EKG), pulse oximetry and non-invasive blood pressure, inhalational induction with 5% sevoflurane was initiated followed by establishment of intravenous (IV) access with 22G cannula. Rocuronium 6 mg with fentanyl 20 mcg was given to facilitate endotracheal intubation. Endotracheal intubation with size 4.5 uncuffed PVC ETT (Euromedical, single use) was accomplished without difficulty. Correct placement of ETT was confirmed by end-tidal CO 2 trace and auscultation. Chest auscultation revealed bilateral but diminished breath sounds and a high resistance was felt at the first manual ventilation. ETT marking at the incisors remained same at 13 cm. The patient was connected to the anesthesia machine, on pressure control mode at an inspiratory pressure of 13 cm H 2 O and a frequency of 20 breaths/minute. The 13 cm H 2 O pressure generated only 30 mL of tidal volume and end-tidal CO 2 showed an obstructive pattern. There was little chest movement during positive pressure ventilation with no gastric insufflation. Breath sounds were equal but distant. An 8 Fr suction catheter was introduced down the lumen of ETT with resistance. No secretions could be suctioned. Two puffs of aerosolized salbutamol were delivered through ETT for assumed bronchospasm, with no improvement. Oxygenation was maintained at 100% with higher ventilatory pressure. The patient was disconnected from the ventilator and the lungs manually ventilated. However, high resistance to manual ventilation continued. The anesthesia circuit was checked for kinks or obstruction but none was found. The ETT marking at the incisors was same at 13 cm.

A possible defect of the ETT was suspected to be the explanation for this clinical scenario. The patient's ETT was removed and substituted with another same sized uncuffed ETT with marked improvement in ventilatory parameters. The surgical procedure was accomplished without any further problems. On closer examination of the removed ETT (Euro medical size 4.5 mm ID, PVC uncuffed), it was found that the lumen of the ETT was almost 50% occluded by a plastic meniscus about 7 mm from the distal end [Figure 1]. The defective ETT was sent to the operating room purchase committee for investigation.


  Discussion Top


After successful endotracheal intubation, ventilation through an ETT may be difficult due to a variety of reasons. The common factors which attribute to difficult ventilation through ETT include anesthesia gas delivery malfunction, obstruction of the breathing circuit (anywhere from common gas outlet to the end of ETT), poor pulmonary compliance (extrinsic or intrinsic), acute severe bronchospasm, tension pneumothorax and endobronchial mass lesion. [3]

A manufacturing defect in an ETT leading to difficulty in ventilation can involve any part of the ETT. Cuff defects leading to herniation of the ETT cuff and intraluminal tracheal obstruction, [4],[5] elliptical defects in the wall of the tube causing air leak, [6] kinking of ETT, [7] and intraluminal plastic films and meniscus [8] causing near total airway obstruction have been described. Fatemah et al. have described a similar case of ETT obstruction from a manufacturing defect in a 7-mm ID PVC cuffed tube due to a plastic film with a small central perforation covering the distal end of the ETT. [9]

Routine preoperative checking of anesthesia equipment - including the ETT - should be a standard of practice. The tube should be checked for patency of the lumen along with the pilot balloon cuff assembly. In spite of the routine preoperative ETT inspection, some manufacturing defects still go unnoticed and become evident only after initiation of mechanical ventilation through that particular ETT. In our case, although the plastic meniscus at the distal end of the tube was significant, it went unrecognized during preoperative inspection. High resistance on manual ventilation and persistently low tidal volume and high inspiratory pressures during pressure control ventilation lead to the suspicion of ETT defect, as all maneuvers to overcome difficulty in ventilation failed. Besides, the absence of a Murphy's eye in older versions of small-sized ETT, as in our case, made ventilation more difficult. A Murphy eye is an essential part of basic ETT design in current practice.

The aim of highlighting this case report is to ensure double checking of ETTs before their use to avoid unnecessary intraoperative problems that can arise from the use of defective ETTs, compromising patient's safety. Most of the ETT defects described in literature could not be identified during routine preoperative inspection. Thus, the role of a vigilant anesthetist is invaluable for patient safety even in the presence of a defective ETT.


  Conclusion Top


Double checking of anesthesia equipment including the ETT, before its use, along with standard monitoring for ventilation, should be a standard of practice for each patient requiring endotracheal intubation.

 
  References Top

1.Gettelman TA, Morris GN. Endotracheal tube failure: undetected by routine testing. Anesth Analg 1995;81:1313.   Back to cited text no. 1  [PUBMED]  [FULLTEXT]  
2.Szekely SM, Webb RK, Williamson JA, Russell WJ. The Australian Incident Monitoring Study: Problems related to the endotracheal tube: An analysis of 2000 incident reports. Anaesth Intensive Care 1993;21:6116.  Back to cited text no. 2      
3.Barst S, Yossefy Y, Lebowitz P. An unusual cause of airway obstruction. Anesth Analg 1994;78:195.   Back to cited text no. 3  [PUBMED]  [FULLTEXT]  
4.Henderson MA. Airway obstruction with a cuffed single-use plastic endotracheal tube. Anaesth Intensive Care 1993;21:370-2.  Back to cited text no. 4  [PUBMED]    
5.Famewo CE. A not so apparent cause of intraluminal tracheal tube obstruction. Anesthesiology 1983;58:593.  Back to cited text no. 5  [PUBMED]  [FULLTEXT]  
6.Lewer BM, Karim Z, Henderson RS. Large air leak from an endotracheal tube due to a manufacturing defect. Anesth Analg 1997;85:944-5.  Back to cited text no. 6  [PUBMED]  [FULLTEXT]  
7.Chua WL, Ng AS. A defective endotracheal tube. Singapore Med J 2002;43:476-8.  Back to cited text no. 7  [PUBMED]    
8.Ngan Kee WD. An unusual problem with an endotracheal tube. Anaesth Intensive Care 1993;21:247-8.  Back to cited text no. 8  [PUBMED]    
9.Hajimohammadi F, Taheri A, Eghtesadi-Araghi P. Obstruction of endotracheal tube: A manufacturing error. Middle East J Anesthesiol 2009;20:303-5.  Back to cited text no. 9  [PUBMED]    


    Figures

  [Figure 1]


This article has been cited by
1 Endotracheal tube connector defect causing airway obstruction in an infant
Divya Sethi,Priyanka Goyal
Southern African Journal of Anaesthesia and Analgesia. 2015; 21(4): 122
[Pubmed] | [DOI]
2 Endotracheal tube connector defect as a cause of high airway pressure
Neerja Bharti,Indu Bala,Kiran Sharma
Pediatric Anesthesia. 2012; 22(5): 502
[Pubmed] | [DOI]



 

Top
 
Previous article Next article
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  IN THIS Article
   Abstract
   Introduction
   Case Report
   Discussion
   Conclusion
   References
   Article Figures

 Article Access Statistics
    Viewed3089    
    Printed189    
    Emailed2    
    PDF Downloaded201    
    Comments [Add]    
    Cited by others 2    

Recommend this journal